Published April 1983 | Version v1
Journal article

Interlaboratory comparison for the quality assurance of 131I-measurements with radionuclide calibrators

Creators

  • 1. Physikalisch-Technische Bundesanstalt, Braunschweig (Germany, F.R.). Lab. fuer Radioaktive Standards

Description

To investigate the accuracy of activity measurements with radionuclide calibrators in nuclear medicine, the Physikalisch-Technische Bundesanstalt (PTB) organized an interinstitute comparison in which 35 hospitals, institutes or laboratories participated. The large majority of the measured results reported agreed with the PTB-value to within +- 10%. This state of affairs is satisfactory. However, major discrepancies may be expected for other radionuclides. This will be studied by future interinstitute comparisons. (orig.)

Abstract (German)

Zur Ueberpruefung der Messsicherheit von Aktivitaetsmessgeraeten in der Nuklearmedizin organisierte die Physikalisch-Technische Bundesanstalt (PTB) einen 131I-Ringversuch, an dem 35 Kliniken, Institute oder Laboratorien teilnahmen. Die gemeldeten Messwerte stimmen - von wenigen Ausnahmen abgesehen - innerhalb von +- 10% mit dem PTB-Messwert ueberein. Dieses Ergebnis ist zufriedenstellend, jedoch muss vermutet werden, dass sich bei anderen Radionukliden groessere Diskrepanzen ergeben. Dies soll bei kuenftigen Ringversuchen untersucht werden. (orig.)

Additional details

Additional titles

Original title (German)
I-Ringversuch zur Ueberpruefung der Messsicherheit von Aktivitaetsmessgeraeten

Publishing Information

Journal Title
Nuc Compact
Journal Volume
14
Journal Issue
2
Series
CODEN: CNNMA.;Nuc Compact.
Journal Page Range
74-78
ISSN
0344-3752