Published February 1989 | Version v1
Journal article

Results of the control of the exposure times of dental X-ray

  • 1. Radiologische Klinik, Gotha (German Democratic Republic)
  • 2. Gesellschaft fuer Stomatologie der DDR

Description

The X-ray gyroscope presented enables the dental radiological personnel to check simply the exposure time of the X-ray unit as to its accuracy. Any load half-wave interval is displayed by a density point on an ORWO XR 5 3/4 x 7 1/2 radiographic film by utilizing the regularity of the emission ratio of one-pulse X-ray generators. A density point corresponds to 0.02 seconds. This method has been used to check and evaluate timers of 10 different X-ray units. The 'Stomax 100' timers showed the least deviations to the exposure time required. (author)

Additional details

Additional titles

Original title (German)
Ergebnisse der Ueberpruefung von Expositionszeiten stomatologischer Roentgendiagnostikeinrichtungen

Publishing Information

Journal Title
Stomatologie der DDR
Journal Volume
39
Journal Issue
2
Series
Stomatol. DDR.
Journal Page Range
123-128
ISSN
0302-4725
CODEN
SMTDA

INIS

Country of Publication
Germany
Country of Input or Organization
German Democratic Republic
INIS RN
20029662
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
BIOMEDICAL RADIOGRAPHY; DENTISTRY; GYROSCOPES; OPTIMIZATION; QUALITY ASSURANCE; RADIATION DOSES; TIME MEASUREMENT; TIMING PROPERTIES; X-RAY EQUIPMENT
Descriptors DEC
DIAGNOSTIC TECHNIQUES; EQUIPMENT; MEDICINE