Published February 1989
| Version v1
Journal article
Results of the control of the exposure times of dental X-ray
Creators
- 1. Radiologische Klinik, Gotha (German Democratic Republic)
- 2. Gesellschaft fuer Stomatologie der DDR
Description
The X-ray gyroscope presented enables the dental radiological personnel to check simply the exposure time of the X-ray unit as to its accuracy. Any load half-wave interval is displayed by a density point on an ORWO XR 5 3/4 x 7 1/2 radiographic film by utilizing the regularity of the emission ratio of one-pulse X-ray generators. A density point corresponds to 0.02 seconds. This method has been used to check and evaluate timers of 10 different X-ray units. The 'Stomax 100' timers showed the least deviations to the exposure time required. (author)
Additional details
Additional titles
- Original title (German)
- Ergebnisse der Ueberpruefung von Expositionszeiten stomatologischer Roentgendiagnostikeinrichtungen
Publishing Information
- Journal Title
- Stomatologie der DDR
- Journal Volume
- 39
- Journal Issue
- 2
- Series
- Stomatol. DDR.
- Journal Page Range
- 123-128
- ISSN
- 0302-4725
- CODEN
- SMTDA
INIS
- Country of Publication
- Germany
- Country of Input or Organization
- German Democratic Republic
- INIS RN
- 20029662
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DENTISTRY; GYROSCOPES; OPTIMIZATION; QUALITY ASSURANCE; RADIATION DOSES; TIME MEASUREMENT; TIMING PROPERTIES; X-RAY EQUIPMENT
- Descriptors DEC
- DIAGNOSTIC TECHNIQUES; EQUIPMENT; MEDICINE