Low dose scanning using 64-detector row computer tomography in pediatric temporal bone
- 1. Department of Radiology, The First Affiliated Hospital of Sun Yat-sen Univ., Guangzhou (China)
Description
Objective: To study the lowest electric current of 64-detector row CT that could manifest the normal structure of pediatric temporal bone. Methods: 180 cases of pediatric temporal bone examined by Toshiba Aquilion 65-detector row CT were collected. These cases were divided into 3 groups (≥150 mA, 30mA, 20mA) randomly. Other scan parameters were fixed. The reconstruct image quality was evaluated by one professor, one associate professor and one technologist randomly with double-blind method. The image quality scanned with routine dose was used as the standard. The image qualities of all the films with different mA were evaluated respectively according to the various structures of temporal bone. All the results were analyzed by Chi-square test. Results: The lowest dose for pediatric temporal bone scanning with 64-detector row CT was 30mA. Conclusion: Low dose CT scanning obtained by 64- detector row can get images that satisfied clinical demand, and it can be used in clinic. (authors)
Additional details
Publishing Information
- Journal Title
- Journal of Diagnostic Imaging and Interventional Radiology
- Journal Volume
- 17
- Journal Issue
- 6
- Journal Page Range
- p. 245-247
- ISSN
- 1005-8001
INIS
- Country of Publication
- China
- Country of Input or Organization
- China
- INIS RN
- 40088103
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CAT SCANNING; CHILDREN; DIAGNOSIS; ELECTRIC CURRENTS; IMAGES; PEDIATRICS; RADIATION DOSES; SKULL
- Descriptors DEC
- AGE GROUPS; ANIMALS; BODY; COMPUTERIZED TOMOGRAPHY; CURRENTS; DIAGNOSTIC TECHNIQUES; DOSES; MAMMALS; MAN; MEDICINE; ORGANS; PRIMATES; SKELETON; TOMOGRAPHY; VERTEBRATES
Optional Information
- Notes