Published October 2004 | Version v1
Journal article

Time resolved neutron flux diagnostics for quasi-steady-state operation study of the HT-7 tokamak

  • 1. Institute of Plasma Physics, Chinese Academy of Sciences, Hefei 230031 (China)

Description

Time resolved neutron flux diagnostic systems based on BF3 proportional counter and ZnS(Ag) scintillator have been developed and implemented on the HT-7 superconducting tokamak. A ten-channel flexible data acquisition system designed with a PCI-8554 general digital counter and industry PC is equipped. Calibrations are made with several neutron sources. The consistencies of experimental data from two techniques have been proven; the BF3 based system is more reliable with better detection efficiency. The measured neutron yield shows good agreement with the simple numerical calculation. The observed photo-neutron production indicates that photon-nuclear reactions are dominant in several special cases such as low density and disruption conditions. Good agreement on ion temperature deduced from neutron diagnosis and neutral particle analyzer under high parameter plasma conditions implies that neutron flux diagnostics can be used as an effective higher temporal resolution ion temperature monitor

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
75
Journal Issue
10
Journal Page Range
p. 4284-4286
ISSN
0034-6748
CODEN
RSINAK

Conference

Title
15. topical conference on high temperature plasma diagnostics
Dates
19-22 Apr 2004
Place
San Diego, CA (United States)

Optional Information

Notes
(c) 2004 American Institute of Physics