A soft x-ray transmission grating spectrometer has been designed for use on high energy-density physics experiments at the National Ignition Facility (NIF); coupled to one of the NIF gated x-ray detectors it records 16 time-gated spectra between 250 and 1000 eV with 100 ps temporal resolution. The trade-off between spectral and spatial resolution leads to an optimized design for measurement of emission around the peak of a 100–300 eV blackbody spectrum. Performance qualification results from the NIF, the Trident Laser Facility and vacuum ultraviolet beamline at the National Synchrotron Light Source, evidence a <100 μm spatial resolution in combination with a source-size limited spectral resolution that is <10 eV at photon energies of 300 eV.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
83
Journal Issue
10
Journal Page Range
p. 10E132-10E132.3
ISSN
0034-6748
CODEN
RSINAK

Optional Information

Notes
(c) 2012 Crown