Development and validation of MCNP4C-based Monte Carlo simulator for fan- and cone-beam x-ray CT
Creators
- 1. Division of Nuclear Medicine, Geneva University Hospital, CH-1211 Geneva 4 (Switzerland)
Description
An x-ray computed tomography (CT) simulator based on the Monte Carlo N-particle radiation transport computer code (MCNP4C) was developed for simulation of both fan- and cone-beam CT scanners. A user-friendly interface running under Matlab 6.5.1 creates the scanner geometry at different views as MCNP4C's input file. The full simulation of x-ray tube, phantom and detectors with single-slice, multi-slice and flat detector configurations was considered. The simulator was validated through comparison with experimental measurements of different nonuniform phantoms with varying sizes on both a clinical and a small-animal CT scanner. There is good agreement between the simulated and measured projections and reconstructed images. Thereafter, the effects of bow-tie filter, phantom size and septa length on scatter distribution in fan-beam CT were studied in detail. The relative difference between detected total, primary and scatter photons for septa length varying between 0 and 95 mm is 11.2%, 1.9% and 84.1%, respectively, whereas the scatter-to-primary ratio decreases by 83.8%. The developed simulator is a powerful tool for evaluating the effect of physical, geometrical and other design parameters on scanner performance and image quality in addition to offering a versatile tool for investigating potential artefacts and correction schemes when using CT-based attenuation correction on dual-modality PET/CT units
Availability note (English)
Available online at http://stacks.iop.org/0031-9155/50/4863/pmb5_20_009.pdf or at the Web site for the journal Physics in Medicine and Biology (ISSN 1361-6560) http://www.iop.org/Additional details
Identifiers
- URL
- http://stacks.iop.org/0031-9155/50/4863/pmb5_20_009.pdf;
- DOI
- 10.1088/0031-9155/50/20/009;
- PII
- S0031-9155(05)99258-0;
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 50
- Journal Issue
- 20
- Journal Page Range
- p. 4863-4885
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 37048662
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ATTENUATION; COMPARATIVE EVALUATIONS; COMPUTER CODES; COMPUTERIZED TOMOGRAPHY; DESIGN; MONTE CARLO METHOD; PERFORMANCE; PHANTOMS; RADIATION TRANSPORT; SIMULATORS; VALIDATION; X-RAY TUBES
- Descriptors DEC
- ANALOG SYSTEMS; CALCULATION METHODS; DIAGNOSTIC TECHNIQUES; ELECTRON TUBES; EQUIPMENT; EVALUATION; FUNCTIONAL MODELS; MOCKUP; STRUCTURAL MODELS; TESTING; TOMOGRAPHY; X-RAY EQUIPMENT