A comparative investigation of Lu2SiO5:Ce and Gd2O2S:Eu powder scintillators for use in x-ray mammography detectors
Creators
- 1. Department of Medical Physics, Medical School, University of Patras, 265 00 Patras (Greece)
- 2. Department of Medical Instruments Technology, Technological Educational Institution of Athens, Egaleo, 122 10 Athens (Greece)
Description
The dominant powder scintillator in most medical imaging modalities for decades has been Gd2O2S:Tb due to the very good intrinsic properties and overall efficiency. Apart from Gd2O2S:Tb, there are alternative powder phosphor scintillators such as Lu2SiO5:Ce and Gd2O2S:Eu that have been suggested for use in various medical imaging modalities. Gd2O2S:Eu emits red light and can be combined mainly with digital mammography detectors such as CCDs. Lu2SiO5:Ce emits blue light and can be combined with blue sensitivity films, photocathodes and some photodiodes. For the purposes of the present study, two scintillating screens, one from Lu2SiO5:Ce and the other from Gd2O2S:Eu powders, were prepared using the method of sedimentation. The screen coating thicknesses were 25.0 and 33.1 mg cm−2 respectively. The screens were investigated by evaluating the following parameters: the output signal, the modulation transfer function, the noise equivalent passband, the informational efficiency, the quantum detection efficiency and the zero-frequency detective quantum efficiency. Furthermore, the spectral compatibility of those materials with various optical detectors was determined. Results were compared to published data for the commercially employed 'Kodak Min-R film-screen system', based on a 31.7 mg cm−2 thick Gd2O2S:Tb phosphor. For Gd2O2S:Eu, MTF data were found comparable to those of Gd2O2S:Tb, while the MTF of Lu2SiO5:Ce was even higher resulting in better spatial resolution and image sharpness properties. On the other hand, Gd2O2S:Eu was found to exhibit higher output signal and zero-frequency detective quantum efficiency than Lu2SiO5:Ce
Availability note (English)
Available from http://dx.doi.org/10.1088/0957-0233/20/10/104008Additional details
Identifiers
- DOI
- 10.1088/0957-0233/20/10/104008;
- PII
- S0957-0233(09)02774-X;
Publishing Information
- Journal Title
- Measurement Science and Technology
- Journal Volume
- 20
- Journal Issue
- 10
- Journal Page Range
- [9 p.]
- ISSN
- 0957-0233
- CODEN
- MSTCEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 45005650
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY; S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- BIOMEDICAL RADIOGRAPHY; DETECTION; EMISSION; FILMS; MAMMARY GLANDS; PHOSPHORS; PHOTOCATHODES; PHOTODIODES; POWDERS; QUANTUM EFFICIENCY; SEDIMENTATION; SENSITIVITY; SIGNALS; SPATIAL RESOLUTION; THICKNESS; TRANSFER FUNCTIONS; X RADIATION
- Descriptors DEC
- BODY; CATHODES; DIAGNOSTIC TECHNIQUES; DIMENSIONS; EFFICIENCY; ELECTRODES; ELECTROMAGNETIC RADIATION; FUNCTIONS; GLANDS; IONIZING RADIATIONS; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIATIONS; RADIOLOGY; RESOLUTION; SEMICONDUCTOR DEVICES; SEMICONDUCTOR DIODES