A thermoluminescence detector (TLD) sensitive to thermal neutrons based on CaF2 (nat) bound with Na2B4O7 has been produced. A second TLD using NaCl as a binder was used to measure the gamma component in the neutron irradiation field. The production techniques were established. Three different grain sizes were used to study the sensitivity and reproducibility of the detectors produced. Tests for neutron fluence determination within the range of interest for radiation protection (1 - 10 x 108 nth.cm-2) showed that the sensitivity of the detectors has to be increased to allow their use in personnel monitoring. (author)

Additional details

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
33
Journal Issue
1-4
Series
Radiat. Prot. Dosim.
Journal Page Range
295-298
ISSN
0144-8420
CODEN
RPDOD

Conference

Title
9. International conference on solid state dosimetry.
Dates
6-10 Nov 1989.
Place
Vienna (Austria).

Optional Information

Notes
Vol. 1 of 2 vols. Vol. 2 published in Radiat. Prot. Dosim. (1990) v. 34(1-4).
Secondary number(s)
CONF--891104.