Published July 1985 | Version v1
Journal article

Polarographic measurement of ascorbate washout in isolated perfused rabbit hearts

  • 1. Univ. of Limburg, Maastricht, Netherland

Description

To study the myocardial washout of ascorbate, the applicability of polarographic detection of ascorbate ions by a platinum electrode (sensitive area 0.03 mm2) was investigated, in both a calibration setup (sampling flow along the electrode: 100 microliter X s-1) and isolated, retrogradely perfused rabbit hearts. In the calibration setup at pH 7.4, the sensitivity of the electrode was 70 microA/mol. This sensitivity increased moderately with increasing pH (13%/unit pH) and increasing sampling flow rate (14% at an increase from 100 to 150 microliter X s-1). In the isolated hearts, ascorbate infused into the aorta was detected in a right ventricular drain by the electrode as well as by the use of 14C-labeled ascorbate. Both recorded time courses were similar except for a scaling factor dependent on flow velocity. During continuous infusion the arteriovenous difference of ascorbate was 2 +/- 2% (SD), indicating a relatively low consumption of ascorbate by the isolated heart. The authors conclude that polarographic measurement of ascorbate in the coronary effluent of an isolated rabbit heart can be performed on-line and relatively easily

Additional details

Publishing Information

Journal Title
Am. J. Physiol.
Journal Issue
p. H150-H154
Series
Am. J. Physiol.
ISSN
0002-9513
CODEN
AJPHA

INIS