Journal article

ACTH-RIA: Comparison of commercially available antisera and standards

  • 1. Freie Univ. Berlin (Germany, F.R.). Inst. fuer Klinische Chemie und Klinische Biochemie

Description

The results demonstrate great differences in sensitivity between the commercial antisera studied. In view of the physiologically low plasma levels of ACTH determination of this peptide using antisera B, C or D is only feasible if ACTH extraction from larger plasma samples precede the final RIA. The great range of sensitivity resulting from the use of different standards with the same antiserum illustrates well the problem of standardization inherent to most protein and peptide RIAs. It strongly indicates the necessity of an uniform standard, especially if external quality control experiments will be put forward. (orig./AJ)

Abstract (German)

Die Ergebnisse zeigeSWINEtersuchten handelsueblichen Sera grosse Unterschiede in der Empfindlichkeit. Angesichts des physiologisch niedrigen Plasma-ACTH-Spiegels ist die Bestimmung dieses Peptids mit Antiserum B, C oder D nur dann moeglich, wenn dem RIA eine ACTH-Extraktion aus groesseren Plasmaproben vorausgeht. Die grossen Empfindlichkeitsschwankungen bei Verwendung verschiedener Standards mit dem gleichen Antiserum verdeutlichen das den meisten Protein- und Peptid-RIAs inhaerente Problem der Standardisierung und beweisen die Notwendigkeit eines einheitlichen Standards, vor allem in Hinblick auf externe Versuche zur Qualitaetskontrolle. (orig./AJ)

Additional details

Publishing Information

Journal Title
Z. Anal. Chem.
Journal Volume
290
Journal Issue
2
Series
Z. Anal. Chem.
Journal Page Range
165-166