Published March 11, 2011 | Version v1
Journal article

Reporting detection efficiency for semiconductor neutron detectors: A need for a standard

  • 1. S.M.A.R.T. Laboratory, Mechanical and Nuclear Engineering Department, Kansas State University, Manhattan, KS 66506 (United States)

Description

In the past few years, there has been an increased interest in the development of alternative neutron detection technologies. Included among many promising alternative neutron detection technologies are semiconductor-based neutron detectors. These detectors are typically small and compact, hence those methods used to characterize large gas-filled neutron detectors are inappropriate. Proposed are standard methods that can be used to model the performance of semiconductor neutron detectors and to characterize experimentally the intrinsic efficiency of these detectors.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2010.12.084

Additional details

Identifiers

DOI
10.1016/j.nima.2010.12.084;
PII
S0168-9002(10)02825-1;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
632
Journal Issue
1
Journal Page Range
p. 167-174
ISSN
0168-9002
CODEN
NIMAER

INIS

Country of Publication
Netherlands
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
42096792
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
EFFICIENCY; NEUTRON DETECTION; NEUTRON DETECTORS; PERFORMANCE; SEMICONDUCTOR DETECTORS; SEMICONDUCTOR MATERIALS
Descriptors DEC
DETECTION; MATERIALS; MEASURING INSTRUMENTS; RADIATION DETECTION; RADIATION DETECTORS

Optional Information

Copyright
Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.