Published October 1992
| Version v1
Journal article
Status of the neutron diagnostic experiment for Alcator C-Mod
- 1. MIT Plasma Fusion Center, Cambridge, Massachusetts 02139 (United States)
Description
The Alcator C-Mod experiment is expected to produce neutrons at a rate of up to 1 x 1016 n/s during its highest performance operation. The global neutron detection system must provide accurate measurement of this production over the full operating regime of the experiment, 1011--1016 n/s. This is accomplished using a series of uranium fission detectors enriched in U235 of staggered sensitivity. These are distributed in two moderator assemblies. A BF3 detector is added to each moderator to cover the lowest operational range. Preliminary calibration of the system has been accomplished using a 65.4 μg Cf252 source placed inside the Alcator C-Mod vacuum vessel. The system design and calibration is discussed herein
Additional details
Publishing Information
- Journal Title
- Review of Scientific Instruments
- Journal Volume
- 63
- Journal Issue
- 10
- Journal Page Range
- p. 4530-4532.
- ISSN
- 0034-6748
- CODEN
- RSINAK
INIS
- Country of Publication
- United States
- Country of Input or Organization
- United States
- INIS RN
- 24017584
- Subject category
- S70: PLASMA PHYSICS AND FUSION TECHNOLOGY; S73: NUCLEAR PHYSICS AND RADIATION PHYSICS;
- Descriptors DEI
- ALCATOR DEVICE; CALIBRATION; CALIFORNIUM 252 TARGET; FISSION NEUTRONS; NEUTRON DETECTION; PLASMA DIAGNOSTICS; URANIUM 235 TARGET
- Descriptors DEC
- BARYONS; CLOSED PLASMA DEVICES; DETECTION; ELEMENTARY PARTICLES; FERMIONS; HADRONS; NEUTRONS; NUCLEONS; RADIATION DETECTION; TARGETS; THERMONUCLEAR DEVICES; TOKAMAK DEVICES