Published August 1988 | Version v1
Journal article

Study on the axial distribution of radioactivity in 60Co source pencils for high dose-rate irradiation and that of gamma-ray intensity

  • 1. Radiation Center of Osaka Prefecture, Sakai (Japan)

Description

Axial radioactivity distributions in 60Co source pencils were measured with a lead collimator and a cavity ionization chamber. Axial γ-ray intensity distributions in the irradiation canister sorrounded by the pencils were measured with a dosimeter consisting of a silicon diode and a charge integrator. The latter distributions well agreed with those calculated from the former distributions by assuming the linear source model. The optimum radioactivity distributions for some arrangement diameters were searched from a point of view of the homogeneity in γ-ray intensity. (author)

Additional details

Publishing Information

Journal Title
Radioisotopes (Tokyo)
Journal Volume
37
Journal Issue
8
Series
Radioisotopes (Tokyo).
Journal Page Range
452-455
ISSN
0033-8303
CODEN
RAISA