On the peculiarities of peak 4 in LiF TLD-100
Creators
- 1. Centro de Investigaciones Energeticas, Medioambientales y Tecnologicas (CIEMAT), Madrid (Spain)
Description
A detailed study of the different glow curves that can be obtained from LiF TLD-100 is presented. This study has been made measuring glow curves after different preparation and readout conditions, and using computerized glow curve analysis to resolve individual glow peaks. In particular, the evolution of peak 4 has been carefully followed. This peak is detected with very different intensity, shape and temperature position depending on the characteristics of the cooling stage of the preparation thermal treatment, and also on the readout heating rate. After very fast cooling and also for fast readout, peak 4 is the most intense peak in the TLD-100 glow curve. The unique behaviour of peak 4 is discussed in the light of the current models for thermoluminescent processes. (Author)
Additional details
Publishing Information
- Journal Title
- Radiation Measurements
- Journal Volume
- 23
- Journal Issue
- 4
- Journal Page Range
- p. 693-701.
- ISSN
- 1350-4487
- CODEN
- RMEAEP
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 27074578
- Subject category
- S36: MATERIALS SCIENCE;
- Resource subtype / Literary indicator
- Numerical Data
- Descriptors DEI
- EXPERIMENTAL DATA; GLOW CURVE; LITHIUM FLUORIDES; THERMOLUMINESCENT DOSEMETERS
- Descriptors DEC
- ALKALI METAL COMPOUNDS; DATA; DOSEMETERS; FLUORIDES; FLUORINE COMPOUNDS; HALIDES; HALOGEN COMPOUNDS; INFORMATION; LITHIUM COMPOUNDS; LITHIUM HALIDES; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; NUMERICAL DATA
Optional Information
- Contract/Grant/Project number
- Contract CEC F13P-CT92-0018