Published December 1972 | Version v1
Journal article

Beam current and position monitor for the Astron accelerator

Description

A beam current and position monitoring system very similar to one developed for the Berkeley ERA accelerator has been developed for use on the Astron accelerator. Beam current measurements are accurate to less than 2%. At beam currents above 500 A beam displacements less than 1 mm off-axis can be observed. The beam current and position at any two of 30 locations in the accelerator and transport section can be simultaneously monitored on either of two oscillograph systems. The current sensors do not interfere with the beam propagation and are outside the vacuum system for easy maintenance and calibration.

Additional details

Identifiers

Publishing Information

Journal Title
Review of Scientific Instruments
Journal Volume
43
Journal Issue
12
Series
Rev. Sci. Instrum.
Journal Page Range
1789-1792
ISSN
0034-6748

INIS

Country of Publication
United States
Country of Input or Organization
United States
INIS RN
4061254
Subject category
S43: PARTICLE ACCELERATORS;
Descriptors DEI
BEAM CURRENTS; BEAM MONITORING; BEAM POSITION; ELECTRON BEAMS; INDUCTION; LINEAR ACCELERATORS
Descriptors DEC
ACCELERATORS; BEAMS; CURRENTS; CYCLIC ACCELERATORS; LEPTON BEAMS; MONITORING; PARTICLE BEAMS

Optional Information

Notes
Updated automatically by Metadata and Full-Text Enrichment Agent