Published November 1985
| Version v1
Journal article
Reusability and fractional read-out in CaF2:Tm (TLD-300)
Creators
- 1. European Organization for Nuclear Research, Geneva (Switzerland)
- 2. N.T.H.U., Taiwan (China)
Description
The first part of this paper is concerned with the reusability of CaF2:Tm (TLD-300) ribbon without thermal treatment between read-out and irradiation. It has been checked up to a dose level of about 200 mGy and the results show that it is possible to reuse this material without annealing in the used dose range. In the second part of the paper a fractional method of TL readings is outlined. The peak separation is checked at different linear testing rates and with different reading times for each peak. The separation technique is then checked in a dose accumulation experiment. (author)
Additional details
Publishing Information
- Journal Title
- Int. J. Appl. Radiat. Isot.
- Journal Volume
- 36
- Journal Issue
- 11
- Series
- Int. J. Appl. Radiat. Isot.
- Journal Page Range
- 893-896
- ISSN
- 0020-708X
- CODEN
- IJARA
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- United Kingdom
- INIS RN
- 17017572
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- ANNEALING; CALCIUM FLUORIDES; MATERIALS TESTING; PEAKS; RADIATION DOSES; READOUT SYSTEMS; THERMOLUMINESCENT DOSEMETERS; THULIUM
- Descriptors DEC
- ALKALINE EARTH METAL COMPOUNDS; CALCIUM COMPOUNDS; CALCIUM HALIDES; DOSEMETERS; ELEMENTS; FLUORIDES; FLUORINE COMPOUNDS; HALIDES; HALOGEN COMPOUNDS; HEAT TREATMENTS; LUMINESCENT DOSEMETERS; MEASURING INSTRUMENTS; METALS; RARE EARTHS; TESTING