Published February 1989 | Version v1
Journal article

Experimental evaluation of a modular scintillation camera for SPECT

  • 1. Div. of Nuclear Medicine, Univ. of Michigan Medical School, Ann Arbor, MI (USA)

Description

A modular scintillation camera has been constructed for use in a ring-type single photon tomograph for Nuclear Medicine. Its performance has been characterized and experiments with coupling the x and y maximum likelihood position estimators and varying the sampling of the scintillation light are described. Average spectral resolution in the x direction is equivalent to a 3.6 mm FWHM Gaussian. In the y direction the response function averages 3.6 mm FWHM. RMS linearity in x is .9 mm and 1.1 mm in y. Potential for improving resolution by increasing the sampling density in x was demonstrated, but a 2-dimensional position estimator would be required to maintain acceptable linearity

Additional details

Publishing Information

Journal Title
IEEE Transactions on Nuclear Science
Journal Volume
36
Journal Issue
1
Series
IEEE Trans. Nucl. Sci.
Journal Page Range
1122-1126
ISSN
0018-9499
CODEN
IETNA

Conference

Title
IEEE nuclear science symposium.
Dates
9-11 Nov 1988.
Place
Orlando, FL (USA).

Optional Information

Secondary number(s)
CONF-881103--.