Published August 2018 | Version v1
Journal article

The efficiency calibration of radioxenon isotopes of inhomogeneous adsorption column for HPGe γ spectrometry

  • 1. Northwest Institute of Nuclear Technology, Xi'an (China)
  • 2. Tsinghua University, Beijing (China). Key Laboratory of Particle and Radiation Imaging of Ministry of Education, Department of Engineering Physics

Description

In this paper, a new efficiency calibration scheme is proposed for the inhomogeneous adsorption column, in which the segment efficiencies modeled by the Monte Carlo toolkit Geant4 are weighted according to the fraction distributions of the adsorbed radioxenon based on the equilibrium adsorption theory. In order to verify the scheme, an experiment of efficiency calibration of HPGe detector by employing radon daughters was performed, and as a result the relative errors between the measured value and simulated value was less than ± 2.3%. This method developed in this paper is useful for the development of rapid test device of radioactive gas. (author)

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
317
Journal Issue
2
Journal Page Range
p. 715-721
ISSN
0236-5731
CODEN
JRNCDM

Optional Information

Notes
17 refs.