Published August 2018
| Version v1
Journal article
The efficiency calibration of radioxenon isotopes of inhomogeneous adsorption column for HPGe γ spectrometry
Creators
- 1. Northwest Institute of Nuclear Technology, Xi'an (China)
- 2. Tsinghua University, Beijing (China). Key Laboratory of Particle and Radiation Imaging of Ministry of Education, Department of Engineering Physics
Description
In this paper, a new efficiency calibration scheme is proposed for the inhomogeneous adsorption column, in which the segment efficiencies modeled by the Monte Carlo toolkit Geant4 are weighted according to the fraction distributions of the adsorbed radioxenon based on the equilibrium adsorption theory. In order to verify the scheme, an experiment of efficiency calibration of HPGe detector by employing radon daughters was performed, and as a result the relative errors between the measured value and simulated value was less than ± 2.3%. This method developed in this paper is useful for the development of rapid test device of radioactive gas. (author)
Additional details
Publishing Information
- Journal Title
- Journal of Radioanalytical and Nuclear Chemistry
- Journal Volume
- 317
- Journal Issue
- 2
- Journal Page Range
- p. 715-721
- ISSN
- 0236-5731
- CODEN
- JRNCDM
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 49085382
- Subject category
- S07: ISOTOPES AND RADIATION SOURCES; S38: RADIATION CHEMISTRY, RADIOCHEMISTRY AND NUCLEAR CHEMISTRY;
- Descriptors DEI
- CALIBRATION; DAUGHTER PRODUCTS; EFFICIENCY; GAMMA SPECTROSCOPY; HIGH-PURITY GE DETECTORS; RADIATION MONITORING; RADON; XENON
- Descriptors DEC
- ELEMENTS; FLUIDS; GASES; GE SEMICONDUCTOR DETECTORS; ISOTOPES; MEASURING INSTRUMENTS; MONITORING; NONMETALS; RADIATION DETECTORS; RARE GASES; SEMICONDUCTOR DETECTORS; SPECTROSCOPY
Optional Information
- Notes
- 17 refs.