Published April 2014 | Version v1
Journal article

A gamma–gamma coincidence/anticoincidence spectrometer for low-level cosmogenic 22Na/7Be activity ratio measurement

Description

In this study, a digital gamma–gamma coincidence/anticoincidence spectrometer was developed and examined for low-level cosmogenic 22Na and 7Be in air-filter sample monitoring. The spectrometer consists of two bismuth germanate scintillators (BGO) and an XIA LLC Digital Gamma Finder (DGF)/Pixie-4 software and card package. The spectrometer design allows a more selective measurement of 22Na with a significant background reduction by gamma–gamma coincidence events processing. Hence, the system provides a more sensitive way to quantify trace amounts of 22Na than normal high resolution gamma spectrometry providing a critical limit of 3 mBq within a 20 h count. The use of a list-mode data acquisition technique enabled simultaneous determination of 22Na and 7Be activity concentrations using a single measurement by coincidence and anticoincidence mode respectively. - Highlights: • A coin/anticoincidence spectrometer to measure 22Na and 7Be by list-mode acquisition. • The method provides a more sensitive way to quantify trace amounts of 22Na. • Compared with HPGe detector, critical limits improved by a factor 6–10. • 22Na and 7Be activity can be determined simultaneously using a single measurement

Availability note (English)

Available from http://dx.doi.org/10.1016/j.jenvrad.2013.12.018

Additional details

Identifiers

DOI
10.1016/j.jenvrad.2013.12.018;
PII
S0265-931X(13)00308-1;

Publishing Information

Journal Title
Journal of Environmental Radioactivity
Journal Volume
130
Journal Page Range
p. 1-6
ISSN
0265-931X
CODEN
JERAEE

Optional Information

Copyright
Copyright (c) 2014 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.