Published August 21, 2011 | Version v1
Journal article

On the energy response function of a CdTe Medipix2 Hexa detector

  • 1. German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg (Germany)
  • 2. Freiburg Materials Research Center (FMF), Stefan-Meier-Strasse 21, 79104 Freiburg (Germany)
  • 3. Erlangen Centre for Astroparticle Physics (ECAP), Erwin-Rommel-Strasse 1, 91058 Erlangen (Germany)

Description

X-ray imaging based on photon counting pixel detectors has received increased interest during the past years. Attached to a semiconductor of choice, some of these devices enable to resolve the spectral components of an image. This work presents the results from measuring the energy response function of a Medipix2 MXR Hexa detector, where six individual Medipix detectors were bump bonded to a 1 mm thick cadmium telluride sensor in order to form a 3x2 array of 4.2x2.8 cm2 size. The average FWHM of the photo peak of an 241Am source was found to be 2.2 and 2.1 keV for single pixels and bias voltages of 200 and 350 V, respectively, across the whole Hexa detector. This corresponds to a relative energy resolution of less than 4%. Adding up all pixel spectra of individual chips lead to an only small deterioration of energy resolution, with line widths of 2.7 and 2.5 keV. In general, a lower detection efficiency was observed for the lower voltage setting, along with a shift of the peak position towards lower energies.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.nima.2010.11.071

Additional details

Identifiers

DOI
10.1016/j.nima.2010.11.071;
PII
S0168-9002(10)02562-3;

Publishing Information

Journal Title
Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
Journal Volume
648
Journal Issue
Suppl.1
Journal Page Range
p. S265-S268
ISSN
0168-9002
CODEN
NIMAER

Conference

Title
4. international conference on imaging techniques in subatomic physics, astrophysics, medicine, biology and industry
Acronym
NIMA
Dates
8-11 Jun 2010
Place
Stockholm (Sweden)

Optional Information

Copyright