A reference air kerma rate calibration protocol for a 370 BGq 192Ir high dose rate brachytherapy source

Creators

  • 1. Austin and Repatriation Medical Centre (ARMC), Heidelberg West (Australia). Medical Engineering and Physics Department

Description

A suggested protocol for the in-air calibration of 192Ir High Dose Rate (HDR) Brachytherapy sources is presented. The protocol was constructed using Nucletron-Oldelft HDR brachytherapy equipment and should be transportable to any High Dose Rate Brachytherapy Afterloader system. A number of correction factors that could be applied for source calibration are recommended for consideration and could be applied if determined as appropriate. (author)

Additional details

Publishing Information

Journal Title
Australasian Physical and Engineering Sciences in Medicine
Journal Volume
20
Journal Issue
2
Journal Page Range
p. 64-70.
ISSN
0158-9938
CODEN
AUPMDI

Optional Information

Notes
19 refs., 3 tabs., 3 figs.