Published April 21, 2016 | Version v1
Journal article

An automated deformable image registration evaluation of confidence tool

  • 1. Department of Radiation Oncology, University of California San Francisco, CA (United States)

Description

Deformable image registration (DIR) is a powerful tool for radiation oncology, but it can produce errors. Beyond this, DIR accuracy is not a fixed quantity and varies on a case-by-case basis. The purpose of this study is to explore the possibility of an automated program to create a patient- and voxel-specific evaluation of DIR accuracy. AUTODIRECT is a software tool that was developed to perform this evaluation for the application of a clinical DIR algorithm to a set of patient images. In brief, AUTODIRECT uses algorithms to generate deformations and applies them to these images (along with processing) to generate sets of test images, with known deformations that are similar to the actual ones and with realistic noise properties. The clinical DIR algorithm is applied to these test image sets (currently 4). From these tests, AUTODIRECT generates spatial and dose uncertainty estimates for each image voxel based on a Student's t distribution. In this study, four commercially available DIR algorithms were used to deform a dose distribution associated with a virtual pelvic phantom image set, and AUTODIRECT was used to generate dose uncertainty estimates for each deformation. The virtual phantom image set has a known ground-truth deformation, so the true dose-warping errors of the DIR algorithms were also known. AUTODIRECT predicted error patterns that closely matched the actual error spatial distribution. On average AUTODIRECT overestimated the magnitude of the dose errors, but tuning the AUTODIRECT algorithms should improve agreement. This proof-of-principle test demonstrates the potential for the AUTODIRECT algorithm as an empirical method to predict DIR errors. (note)

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/61/8/N203

Additional details

Identifiers

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
61
Journal Issue
8
Journal Page Range
p. N203-N214
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47107394
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; ALGORITHMS; COMPUTER CODES; DEFORMATION; ERRORS; EVALUATION; GROUND TRUTH MEASUREMENTS; IMAGES; NOISE; PATIENTS; PHANTOMS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; SPATIAL DISTRIBUTION
Descriptors DEC
DISTRIBUTION; DOSES; MATHEMATICAL LOGIC; MOCKUP; STRUCTURAL MODELS