Published February 1988
| Version v1
Journal article
New development of MRI lung nodule simulator for detecting on pulmonary nodule
Creators
- 1. Kobe Univ. (Japan). School of Medicine
Description
In an effort to augment the value of MRI in detecting lung nodules, we devised a new lung nodule phantom made of polyvinylalcohol. Besides, a thoracic phantom made of polyvinyl alcohol, solidoils, and artificial bone was prepared to evaluate the environmental effect around the nodule phantom. The p, T1 and T2 values of these simulators were equivalent to those of chest wall tissues. T1 and T2 of these phantoms were calculated by use of a 0.2T resistive magnetic resonance machine. Our findings suggested that T1 and T2 of the lung nodule phantom varies as a function of position. (author)
Additional details
Additional titles
- Subtitle (English)
- Experimental study
Publishing Information
- Journal Title
- Nippon Igaku Hoshasen Gakkai Zasshi
- Journal Volume
- 48
- Journal Issue
- 2
- Series
- Nippon Igaku Hoshasen Gakkai Zasshi.
- Journal Page Range
- 212-213
- ISSN
- 0048-0428
- CODEN
- NHGZA
INIS
- Country of Publication
- Japan
- Country of Input or Organization
- Japan
- INIS RN
- 19076925
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- CARCINOMAS; CHEST; IMAGES; LUNGS; NMR IMAGING; PHANTOMS; SIMULATORS; SPIN-LATTICE RELAXATION; SPIN-SPIN RELAXATION
- Descriptors DEC
- ANALOG SYSTEMS; BODY; BODY AREAS; DIAGNOSTIC TECHNIQUES; DISEASES; FUNCTIONAL MODELS; MOCKUP; NEOPLASMS; ORGANS; RELAXATION; RESPIRATORY SYSTEM; STRUCTURAL MODELS