Published 1984 | Version v1
Book

Fast harmonic field mapper

  • 1. National Superconducting Cyclotron Laboratory, Michigan State University, East Lansing, Michigan

Description

In early 1983 it was decided to mount coils on arms separated by 120 degrees and buck them out so that the third harmonic dphi/dt component would be cancelled and thus the first and second field harmonics could be very accurately measured. The original intention was to do as others had done, namely, use fast ADC's to read the voltages, and computer process the result to get the Fourier components. However, because of the 100 to 1 dynamic range of the fast ADC's and the likelihood that noise would be a problem, the authors decided to do things differently. Using a fast Fourier transform analyzer was considered, but this instrument is very expensive, so they decided to use a completely electronic analog approach: The authors decided to use active bandpass filters to render the harmonic components

Additional details

Publishing Information

Publisher
Institution of Electrical and Electronics Engineers.
Imprint Place
New York, NY (USA)
Imprint Title
Tenth international conference on cyclotrons and their applications
Journal Page Range
p. 104-106.

Conference

Title
10. international conference on cyclotrons and their applications.
Dates
30 Apr - 3 May 1984.
Place
East Lansing, MI (USA).