Detection and dosimetry of gamma ray emitted from thallium-201 and technetium-99m based on chemiluminescence technique

  • 1. Laboratory of Microanalysis, Institute of Biochemistry and Biophysics, University of Tehran, P.O. Box 13145-1384, Tehran (Iran, Islamic Republic of)
  • 2. Faculty of Medicine, Department of Physiology and Biophysics, Faculty of Medicine, Baqiyatollah University of Medical Sciences, P.O. Box 19395-6558, Tehran (Iran, Islamic Republic of)

Description

This report describes the detection and dosimetry of gamma ray emitted from Thallium-201 (201Tl) and Technetium-99m (99mTc) based on chemiluminescence technique. H2O2 produced by two gamma emitter radioisotopes of 201Tl and 99mTc were quantitatively measured by chemiluminescence method. Upon producing H2O2 in a luminol alkaline solution, in the presence of diperiodatocuprate, as catalyst a chemical reaction was accrued and consequently the emitted light was measured. The determined H2O2 concentration was correlated with the gamma ray detection and dosimetry. The sensitivity of chemiluminescence technique for 201Tl and 99mTc dosimetry was determined to be 0.20 and 0.08 MBq/l (Mega Becquerel per liter) respectively (R.S.D. = %5, N = 3). The plotted calibration curves showed detection limits of 3.24 and 1.76 MBq/l for 201Tl and 99mTc, respectively.

Availability note (English)

Available from http://dx.doi.org/10.1016/j.radmeas.2010.06.002

Additional details

Identifiers

DOI
10.1016/j.radmeas.2010.06.002;
PII
S1350-4487(10)00193-9;

Publishing Information

Journal Title
Radiation Measurements
Journal Volume
45
Journal Issue
8
Journal Page Range
p. 906-910
ISSN
1350-4487
CODEN
RMEAEP

Optional Information

Copyright
Copyright (c) 2010 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.