Published November 15, 1976
| Version v1
Journal article
Specific radioactivity of radioimmunoassay tracer determined by self-displacement: a re-evaluation
Description
An error was discovered in the method of self-displacement commonly used for determining the specific radioactivity of labelling tracer antigen used in radioimmunoassay; viz. that the calculation does not take into account the mass of labelled antigen present in the mixture containing unlabelled antigen standard. A simple graphical plot is described to obviate this error so that the correct value for the specific radioactivity of a radioimmunoassay tracer can be reliably estimated
Additional details
Identifiers
Publishing Information
- Journal Title
- Clinica Chimica Acta
- Journal Volume
- 73
- Journal Issue
- 1
- Series
- Clin. Chim. Acta.
- Journal Page Range
- 213-216
- ISSN
- 0009-8981
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- Netherlands
- INIS RN
- 8332576
- Subject category
- S60: APPLIED LIFE SCIENCES;
- Descriptors DEI
- ANTIGEN-ANTIBODY REACTIONS; BIOCHEMICAL REACTION KINETICS; ISOTOPE DILUTION; RADIOACTIVITY; RADIOIMMUNOASSAY
- Descriptors DEC
- ISOTOPE APPLICATIONS; KINETICS; REACTION KINETICS; TRACER TECHNIQUES
Optional Information
- Notes
- Updated automatically by Metadata and Full-Text Enrichment Agent