Performance evaluation of Siemens E.CAM(R) SPECT system at Korle-Bu Teaching Hospital (Ghana)
Creators
- 1. Radiological and Medical Science Research Institute, Ghana Atomic Energy Commission, Accra (Ghana)
- 2. School of Nuclear and Allied Sciences, University of Ghana, Atomic Campus, Accra (Ghana)
Description
Performance evaluation tests on Siemens e.cam(R) Signature Series Single Photon Emission Computed Tomography (SPECT)System at Korle-Bu Teaching Hospital (Ghana) were conducted for compliance with Manufacturer's specifications and user's requirements. The tests were performed according to National Electrical Manufacturer's Association protocols, and different measuring instruments and phantoms containing 99m-Tc were used. Extrinsic uniformity, system energy resolution, system spatial resolution (without scattering), detector shield leakage and system planar sensitivity were evaluated. The system energy resolution measured was 9.38 % as against manufacturer's specification of ≤ 9.9 %, Planar sensitivity value measured was 232.92 c/min/mCi while manufacturers' specification was 202 c/min/mCi, and detector shield leakage values measured were 0.001 %, 0.002 % and 0.025 % against an accepted value of < 1 %. The performance evaluation confirmed that the system met requirements for clinical nuclear medicine imaging, and the values could be used as database for future quality control measurements. (au)
Additional details
Publishing Information
- Journal Title
- Journal of Applied Science and Technology
- Journal Volume
- 16
- Journal Issue
- 1-2
- Journal Page Range
- p. 121-126
- ISSN
- 0855-2215
INIS
- Country of Publication
- Ghana
- Country of Input or Organization
- Ghana
- INIS RN
- 42066883
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Descriptors DEI
- ENERGY RESOLUTION; EVALUATION; GAMMA CAMERAS; NAI DETECTORS; NUCLEAR MEDICINE; PERFORMANCE TESTING; QUALITY CONTROL; SHIELDING; SINGLE PHOTON EMISSION COMPUTED TOMOGRAPHY; SPATIAL RESOLUTION; TECHNETIUM 99
- Descriptors DEC
- BETA DECAY RADIOISOTOPES; BETA-MINUS DECAY RADIOISOTOPES; CAMERAS; COMPUTERIZED TOMOGRAPHY; CONTROL; DIAGNOSTIC TECHNIQUES; EMISSION COMPUTED TOMOGRAPHY; HOURS LIVING RADIOISOTOPES; INTERMEDIATE MASS NUCLEI; INTERNAL CONVERSION RADIOISOTOPES; ISOMERIC TRANSITION ISOTOPES; ISOTOPES; MEASURING INSTRUMENTS; MEDICINE; NUCLEI; ODD-EVEN NUCLEI; RADIATION DETECTORS; RADIOISOTOPES; RESOLUTION; SCINTILLATION COUNTERS; SOLID SCINTILLATION DETECTORS; TECHNETIUM ISOTOPES; TESTING; TOMOGRAPHY; YEARS LIVING RADIOISOTOPES
Optional Information
- Notes
- 8 figs., 2 tabs., 17 refs.