Radiation exposures to technologists from nuclear medicine imaging procedures

  • 1. Pasqua Hospital, Regina, Saskatchewan (Canada)

Description

Radiation exposures incurred by nuclear medicine technologists during diagnostic imaging and gamma camera quality control (QC) were measured on a procedural basis over a three-month period using a portable, low-range, self-reading ion chamber. A total of more than 400 measurements were made for 15 selected procedures. From these, mean procedural exposures and standard deviations were calculated. The results show that daily flood phantom QC, at 0.58 mR, and gated cardiac studies, at 0.45 mR, were the two greatest sources of exposure. Other procedures resulted in exposures varying roughly from 0.10 to 0.20 mR. Difficult patients were responsible for a doubling of technologist exposure for many procedures. Standard deviations were large for all procedures, averaging 65% of the mean values. Comparison of technologist exposure inferred from the procedural measurements with the time coincident collective dose equivalent recorded by the TLD service of the Radiation Protection Bureau indicates that approximately half of the collective technologist exposure arose from patient handling and flood QC