Published 1983 | Version v1
Journal article

Testing of camera performance standards at steady and local overloading

  • 1. Gamma Muevek, Budapest (Hungary)
  • 2. Jozsef Attila Tudomanyegyetem, Szeged (Hungary)

Description

Camera performance standards are usually given for low count rates and uniform irradiation. A conventional analog gamma camera system (Gamma MB 9100 manufactured under the know-how of Picker DC 4/12 in Hungary) was studied on the basis of the Picker test procedure and the NEMA standard system. Uniformity, linearity, spatial and energy resolution were measured at high count rates and uniform irradiation (steady overloading). Linearity was studied at local overloading. Linearity, spatial and energy resolution were measured by a 1024-channel analyzer-computer system of KFKI, Hungary. The data were evaluated on the basis of NEMA standards, while uniformity was measured by the Gamma data processing system and evaluated by special SEGAMS programs. Performance variations were studied between 7500 cps and 75.000 cps pulse rates. Spatial and energy resolution were influenced strongly, uniformity slightly by pulse rates, while linearity did not change at all. Linearity was not influenced even by local overloading. (author)

Additional details

Additional titles

Original title (Hungarian)
Gammakamera specifikacios parameterei egyenletes es lokalis impulzusterhelesnel

Publishing Information

Journal Title
Izotoptechnika
Journal Volume
26
Journal Issue
2
Series
Izotoptechnika.
Journal Page Range
133-143
ISSN
0004-7201

INIS

Country of Publication
Hungary
Country of Input or Organization
Hungary
INIS RN
15007032
Subject category
S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
Descriptors DEI
COUNTING RATES; ENERGY RESOLUTION; GAMMA CAMERAS; IMAGES; NONUNIFORM IRRADIATION; PERFORMANCE TESTING; SPATIAL RESOLUTION
Descriptors DEC
CAMERAS; IRRADIATION; RESOLUTION; TESTING

Optional Information

Notes
3 refs.; 12 figs.