Testing of camera performance standards at steady and local overloading
- 1. Gamma Muevek, Budapest (Hungary)
- 2. Jozsef Attila Tudomanyegyetem, Szeged (Hungary)
Description
Camera performance standards are usually given for low count rates and uniform irradiation. A conventional analog gamma camera system (Gamma MB 9100 manufactured under the know-how of Picker DC 4/12 in Hungary) was studied on the basis of the Picker test procedure and the NEMA standard system. Uniformity, linearity, spatial and energy resolution were measured at high count rates and uniform irradiation (steady overloading). Linearity was studied at local overloading. Linearity, spatial and energy resolution were measured by a 1024-channel analyzer-computer system of KFKI, Hungary. The data were evaluated on the basis of NEMA standards, while uniformity was measured by the Gamma data processing system and evaluated by special SEGAMS programs. Performance variations were studied between 7500 cps and 75.000 cps pulse rates. Spatial and energy resolution were influenced strongly, uniformity slightly by pulse rates, while linearity did not change at all. Linearity was not influenced even by local overloading. (author)
Additional details
Additional titles
- Original title (Hungarian)
- Gammakamera specifikacios parameterei egyenletes es lokalis impulzusterhelesnel
Publishing Information
- Journal Title
- Izotoptechnika
- Journal Volume
- 26
- Journal Issue
- 2
- Series
- Izotoptechnika.
- Journal Page Range
- 133-143
- ISSN
- 0004-7201
INIS
- Country of Publication
- Hungary
- Country of Input or Organization
- Hungary
- INIS RN
- 15007032
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- COUNTING RATES; ENERGY RESOLUTION; GAMMA CAMERAS; IMAGES; NONUNIFORM IRRADIATION; PERFORMANCE TESTING; SPATIAL RESOLUTION
- Descriptors DEC
- CAMERAS; IRRADIATION; RESOLUTION; TESTING
Optional Information
- Notes
- 3 refs.; 12 figs.