Dose-rate to water calibrations for brachytherapy sources from the end-user perspective
Creators
- 1. Clinic of Radiotherapy, Campus Kiel, UK S-H, Arnold-Heller-Str. 3, Haus 23, (Germany)
- 2. Department of Medical Physics, Instituut Verbeeten, Tilburg, (Netherlands)
- 3. Department of Medical Physics, Division of Surgery and Cancer Medicine, Oslo University Hospital -The Norwegian Radium Hospital, Oslo, (Norway)
- 4. Medical Physics Laboratory, Medical School, University of Athens, Athens, (Greece)
- 5. Department of Radiation Oncology, Europe Hospitals, Brussels, (Belgium)
- 6. Department of Radiation Oncology, Tufts University School of Medicine, Boston, MA, (United States)
Description
Independent primary standards for brachytherapy photon-emitting source calibration in terms of dose-rate to water have been developed within the framework of the Euramet T2.J06 project. The introduction of dose-rate to water calibration presents an important change in clinical brachytherapy dosimetry that is expected to result to improved dosimetric accuracy. Nevertheless, as with any change in dosimetry for radiation therapy purposes, a phase-in period of well concerted actions aimed at precluding ambiguities and accidents at the end-user level is necessary. The overall uncertainty budget of clinical brachytherapy applications, as well as current trends in brachytherapy treatment planning system dose-calculation algorithms, also need to be considered for a realistic assessment of the net benefit of improving source calibration accuracy. (authors)
Availability note (English)
Available from doi: http://dx.doi.org/10.1088/0026-1394/49/5/S249Additional details
Identifiers
Publishing Information
- Journal Title
- Metrologia
- Journal Volume
- 49
- Journal Issue
- no.5
- Journal Page Range
- p. S249-S252
- ISSN
- 0026-1394
- CODEN
- MTRGAU
Conference
- Title
- Conference on Advanced Metrology for Cancer Therapy 2011
- Dates
- 29 Nov - 1 Dec 2011
- Place
- Braunschweig (Germany)
INIS
- Country of Publication
- France
- Country of Input or Organization
- France
- INIS RN
- 44026726
- Subject category
- S61: RADIATION PROTECTION AND DOSIMETRY; S62: RADIOLOGY AND NUCLEAR MEDICINE;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ACCURACY; BRACHYTHERAPY; CALIBRATION STANDARDS; DATA COVARIANCES; DOSE RATES; SPECIFICATIONS; WATER
- Descriptors DEC
- HYDROGEN COMPOUNDS; MEDICINE; NUCLEAR MEDICINE; OXYGEN COMPOUNDS; RADIOLOGY; RADIOTHERAPY; STANDARDS; THERAPY
Optional Information
- Notes
- 19 refs.