Published March 1, 2016 | Version v1
Journal article

The systematic and random errors determination using realtime 3D surface tracking system in breast cancer

  • 1. Medical Physics School, Faculty of Medicine, Ramathibodi Hospital, Mahidol University, Bangkok (Thailand)
  • 2. Department of Radiology, King Chulalongkorn Memorial Hospital, The Thai Red Cross Society, Bangkok (Thailand)
  • 3. Department of Radiological Technology, Faculty of Medical Technology, Mahidol University, Bangkok (Thailand)

Description

The purpose of this study to determine the patient setup uncertainties in deep inspiration breath-hold (DIBH) radiation therapy for left breast cancer patients using real-time 3D surface tracking system. The six breast cancer patients treated by 6 MV photon beams from TrueBeam linear accelerator were selected. The patient setup errors and motion during treatment were observed and calculated for interfraction and intrafraction motions. The systematic and random errors were calculated in vertical, longitudinal and lateral directions. From 180 images tracking before and during treatment, the maximum systematic error of interfraction and intrafraction motions were 0.56 mm and 0.23 mm, the maximum random error of interfraction and intrafraction motions were 1.18 mm and 0.53 mm, respectively. The interfraction was more pronounce than the intrafraction, while the systematic error was less impact than random error. In conclusion the intrafraction motion error from patient setup uncertainty is about half of interfraction motion error, which is less impact due to the stability in organ movement from DIBH. The systematic reproducibility is also half of random error because of the high efficiency of modern linac machine that can reduce the systematic uncertainty effectively, while the random errors is uncontrollable. (paper)

Availability note (English)

Available from http://dx.doi.org/10.1088/1742-6596/694/1/012005

Additional details

Publishing Information

Journal Title
Journal of Physics. Conference Series (Online)
Journal Volume
694
Journal Issue
1
Journal Page Range
[5 p.]
ISSN
1742-6596

Conference

Title
13. South-East Asian congress of medical physics (SEACOMP) 2015
Dates
10-12 Dec 2015
Place
Yogyakarta (Indonesia)

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
47117942
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Resource subtype / Literary indicator
Conference
Descriptors DEI
BREATH; EFFICIENCY; ERRORS; IMAGES; LINEAR ACCELERATORS; MAMMARY GLANDS; NEOPLASMS; PATIENTS; PHOTON BEAMS; PHOTONS; RADIOTHERAPY; RANDOMNESS; STABILITY; SURFACES
Descriptors DEC
ACCELERATORS; BEAMS; BODY; BOSONS; DISEASES; ELEMENTARY PARTICLES; GLANDS; MASSLESS PARTICLES; MEDICINE; NUCLEAR MEDICINE; ORGANS; RADIOLOGY; THERAPY