Published September 2012 | Version v1
Journal article

Assessment of accuracy and efficiency of atlas-based autosegmentation for prostate radiotherapy in a variety of clinical conditions

  • 1. Medical Univ. Vienna/AKH Vienna (Austria). Div. of Medical Radiation Physics
  • 2. Medical Univ. Vienna (Austria). Christian Doppler Lab. for Medical Radiation Research for Radiation Oncology
  • 3. Medical Univ. Vienna (Austria). Center for Medical Physics and Biomedical Engineering

Description

Background and purpose: The goal of the current study was to evaluate the commercially available atlas-based autosegmentation software for clinical use in prostate radiotherapy. The accuracy was benchmarked against interobserver variability. Material and methods: A total of 20 planning computed tomographs (CTs) and 10 cone-beam CTs (CBCTs) were selected for prostate, rectum, and bladder delineation. The images varied regarding to individual (age, body mass index) and setup parameters (contrast agent, rectal balloon, implanted markers). Automatically created contours with ABAS registered and iPlan registered were compared to an expert's delineation by calculating the Dice similarity coefficient (DSC) and conformity index. Results: Demo-atlases of both systems showed different results for bladder (DSCABAS 0.86 ± 0.17, DSCiPlan 0.51 ± 0.30) and prostate (DSCABAS 0.71 ± 0.14, DSCiPlan 0.57 ± 0.19). Rectum delineation (DSCABAS 0.78 ± 0.11, DSCiPlan 0.84 ± 0.08) demonstrated differences between the systems but better correlation of the automatically drawn volumes. ABAS registered was closest to the interobserver benchmark. Autosegmentation with iPlan registered, ABAS registered and manual segmentation took 0.5, 4 and 15-20 min, respectively. Automatic contouring on CBCT showed high dependence on image quality (DSC bladder 0.54, rectum 0.42, prostate 0.34). Conclusion: For clinical routine, efforts are still necessary to either redesign algorithms implemented in autosegmentation or to optimize image quality for CBCT to guarantee required accuracy and time savings for adaptive radiotherapy. (orig.)

Additional details

Publishing Information

Journal Title
Strahlentherapie und Onkologie
Journal Volume
188
Journal Issue
9
Journal Page Range
p. 807-813
ISSN
0179-7158
CODEN
STONE4