Published December 1995 | Version v1
Journal article

A miniaturized rapid paper chromatographic procedure for quality control of technetium-99m sestamibi

  • 1. Kuwait Central Radiopharmacy and Dept. of Nuclear Medicine, Kuwait Cancer Control Center (Kuwait)
  • 2. Kuwait Univ. (Kuwait). Faculty of Medicine

Description

A miniaturized rapid paper chromatographic technique (MRPC) for quality control of a technetium-99m sestamibi preparation was developed and compared with the manufacturer's and Hung et al. techniques. The MRPC system involves the use of 6.0x0.5 cm Whatman 3MM paper strip developed in ethyl acetate. The procedure was completed within 3 min while that of the manufacturer and Hung techniques took 30-35 and 4 min respectively. The Rf range of 99mTc-sestamibi using MRPC was 0.55-0.75 while that of the other two techniques was 0.9-1.0. The results indicate that MRPC can be used to separate 99mTc-sestamibi from any 99mTc contaminant that migrates with the solvent front. The MRPC is a fast and effective chromatographic technique for routine quality control testing of 99mTc-sestamibi preparation. (orig.)

Additional details

Publishing Information

Journal Title
European Journal of Nuclear Medicine
Journal Volume
22
Journal Issue
12
Journal Page Range
p. 1416-1419.
ISSN
0340-6997
CODEN
EJNMD9