Published February 2014 | Version v1
Journal article

Calibration of the 90Sr+90Y ophthalmic and dermatological applicators with an extrapolation ionization minichamber

  • 1. Instituto de Pesquisas Energéticas e Nucleares, Comissão Nacional de Energia Nuclear, IPEN/CNEN-SP, Av. Prof. Lineu Prestes, 2242, 05508-000 São Paulo, SP (Brazil)
  • 2. Centro Regional de Ciências Nucleares, Comissão Nacional de Energia Nuclear, CRCN/CNEN, Av. Prof. Luis Freire, 200, 50740-540 Recife, PE (Brazil)

Description

90Sr+90Y clinical applicators are used for brachytherapy in Brazilian clinics even though they are not manufactured anymore. Such sources must be calibrated periodically, and one of the calibration methods in use is ionometry with extrapolation ionization chambers. 90Sr+90Y clinical applicators were calibrated using an extrapolation minichamber developed at the Calibration Laboratory at IPEN. The obtained results agree satisfactorily with the data provided in calibration certificates of the sources. - Highlights: • 90Sr+90Y clinical applicators were calibrated using a mini-extrapolation chamber. • An extrapolation curve was obtained for each applicator during its calibration. • The results were compared with those provided by the calibration certificates. • All results of the dermatological applicators presented lower differences than 5%

Availability note (English)

Available from http://dx.doi.org/10.1016/j.apradiso.2013.10.022

Additional details

Identifiers

DOI
10.1016/j.apradiso.2013.10.022;
PII
S0969-8043(13)00417-X;

Publishing Information

Journal Title
Applied Radiation and Isotopes
Journal Volume
85
Journal Page Range
p. 11-13
ISSN
0969-8043
CODEN
ARISEF

Optional Information

Copyright
Copyright (c) 2013 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.