Published 2020 | Version v1
Journal article

Effect of secondary radiation from lead shielding during liquid scintillation counting

Creators

  • 1. National Institute of Advanced Industrial Science and Technology (AIST), National Metrology Institute of Japan, Tsukuba, Ibaraki (Japan)

Description

In this study, the effects of secondary radiation from the lead shielding on the efficiency calculation were estimated, using the MICELLE2 program and the EGS5MPI Monte Carlo code. The results show that the counting efficiency values for 54Mn (which decays via electron capture and emits gamma rays) were increased by approximately 0.5% due to the effect of the lead shield. Thus, secondary radiation was found to be non-negligible during the standardization of this radionuclide, leading to an standard uncertainty in the range of 0.6-1.2%. Henceforth, secondary radiation should be taken into account during the standardization of nuclides that decay by electron capture and emit gamma rays when using liquid scintillation counters. (author)

Availability note (English)

Available from DOI: https://doi.org/10.3769/radioisotopes.69.117

Additional details

Identifiers

Publishing Information

Journal Title
Radioisotopes (Online)
Journal Volume
69
Journal Issue
3
Journal Page Range
p. 117-120
ISSN
1884-4111

Optional Information

Notes
8 refs., 3 figs., 1 tab.