Comparison of a continuous working level monitor for radon daughters with conventional grab-sampling

Description

An evaluation of a radon daughter monitor was carried out under laboratory controlled conditions. The monitor operates on continuous sampling and time integrating principles and was tested in conjunction with a newly designed, large radon/thoron room calibration facility. The monitor was tested under constant and rapidly fluctuating radiation conditions. Experimental data obtained with the monitor were compared with data obtained by conventional grab-sampling and with an automated radon daughter/thoron daughter 'grab-sampler'. The Working Level used in the tests ranged from less than 0.01 WL to approximately 10 WL. The measurements were carried out under low aerosol concentration (1 x 103 - 2 x 103 cm-3, approximately) to study plate-out effects in the sampling head. Good agreement (within about 10 %) was found between the monitor, conventional grab-sampling and the automated grab-sampler. The monitor should prove quite flexible, useful and reliable for monitoring underground and surface environments in the uranium mining industry

Availability note (English)

MF available from INIS under the Report Number.

Files

16011129.pdf

Files (280.8 kB)

Name Size Download all
md5:2eb7cbd3380c3e7287c56881615fa31d
280.8 kB Preview Download

Additional details

Publishing Information

Imprint Pagination
12 p.
Report number
MRP/MRL--82-79