Dose to population as a metric in the design of optimised exposure control in digital mammography

  • 1. Mammography Engineering Dept., GE Healthcare, 283 rue de la Miniere, 78530 Buc (France)

Description

This paper describes a methods for automatic optimisation of parameter (AOP) in digital mammography systems. Using a model of the image chain, contrast to noise ratio (CNR) and average glandular dose (AGD) are computed for possible X-ray parameters and breast types. The optimisation process consists of the determination of the operating points providing the lowest possible AGD for each CNR level and breast type. The proposed metric for the dose used in the design of an AOP mode is the resulting dose to the population, computed by averaging the AGD values over the distribution of breast types in the population. This method has been applied to the automatic exposure control of new digital mammography equipment. Breast thickness and composition are estimated from a low dose pre-exposure and used to index tables containing sets of optimised operating points. The resulting average dose to the population ranges from a level comparable to state-of-the-art screen/film mammography to a reduction by a factor of two. Using this method, both CNR and dose are kept under control for all breast types, taking into consideration both individual and collective risk. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/nch579

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
114
Journal Issue
1-3
Journal Page Range
p. 369-374
ISSN
0144-8420

Conference

Title
2. Malmoe Conference on Medical X-Ray Imaging - Optimisation strategies in medical X-ray imaging
Dates
23-25 Apr 2004
Place
Malmoe (Sweden)

Optional Information

Notes
20 refs