Reduced dose and improved image quality with computerized line-scan radiography system

  • 1. Pittsburgh Univ., PA (United States)
  • 2. Scantech Corp., Columbus, OH (USA)

Description

This paper reports on Diode-Array Digital Radiography (DADR) that reduces patient x-ray dose and improves diagnostic quality compared with conventional filmed radiography. DADR geometrically eliminates scattered x-rays by using a line-scanning technique, removing background fog. The x-ray tube is collimated to produce a fan beam of x-rays which impinges on a linear detector. The source and detector are held in alignment by a C-arm which scans the patient by translating the assembly orthogonally to the detector aperture. Each image line is exposed for a very brief time, which provides for a very short effective x-ray exposure and eliminates blurring due to patient motion. The detector uses a patented combination of scintillator, fiber optic image guides and self-scanning photodiode arrays which, in combination with scatter rejection, creates an electronic image having greater dynamic range than filmed x-rays. The image is converted to 12-bit digital computer data which may be sorted inexpensively, enhanced for diagnosis, and transferred to a PACS network. Images are displayed in full 2352 x 2000 resolution on a high-bandwidth monitor. Diagnostic experiments have concentrated on chest imaging, an application for which Scantech is developing a new product. Patient dose is typically less than half that incurred with conventional techniques

Additional details

Publishing Information

Publisher
IEEE Service Center.
Imprint Place
Piscataway, NJ (USA)
Imprint Title
IEEE nuclear science symposium conference record emdash 1990
Imprint Pagination
1636 p.
Journal Page Range
p. 1102-1118.

Conference

Title
1990 Institute of Electrical and Electronics Engineers (IEEE) nuclear science symposium.
Dates
22-27 Oct 1990.
Place
Arlington, VA (USA).

Optional Information

Secondary number(s)
CONF-9010220--.