A novel silicon detector for neutral particle analysis in JET fusion research
Creators
- 1. VTT, Microelectronics, Tietotie 3, Espoo, P.O. Box 1000, FI-02044 VTT (Finland)
- 2. Helsinki Institute of Physics, High Energy Physics, University of Helsinki, P.O. Box 64, FI-00014 Helsinki (Finland)
- 3. Helsinki University of Technology, Association Euratom-TEKES, P.O. Box 4100, FI-02015 TKK (Finland)
Description
A novel detector for the neutral particle analyzers (NPA) of the Joint European Torus (JET) is introduced for studying plasma characteristics during the fusion experiments. The proposed ion detector for the NPA would replace the presently used very thin scintillator-photomultiplier tube combination. The scintillator detectors exhibit poor energy resolution, their non-linearity precludes ion identification and their sensitivity to background makes signal-to-background separation difficult under the severe radiation background present during JET plasma pulses. The proposed new NPA detector is based on direct conversion of charge in silicon and approximate matching of the detector thickness with the ranges of the observed particles. The light ions of interest to the NPA have ranges that vary from less than 1 to 25 μm. A thin silicon detector is only weakly sensitive to photon and neutron backgrounds but detects highly ionizing ions efficiently. Even high-energy gammas deposit only a little energy in the thin detector allowing effective background discrimination through the pulse-height analysis. Thin silicon strip detectors have been fabricated by using the silicon-on-insulator (SOI) technology. Fabricated detectors have 6 or 26 μm thick high resistive silicon bonded on a conductive silicon support. The thinner detectors are designed to be used for the low energy NPA and the thicker ones for the high-energy NPA. The paper comprises an introduction of the fabricated detector structures, results of the preliminary electrical characterization and the first spectroscopic measurements
Availability note (English)
Available from http://dx.doi.org/10.1016/j.nima.2008.03.030Additional details
Identifiers
- DOI
- 10.1016/j.nima.2008.03.030;
- PII
- S0168-9002(08)00409-9;
Publishing Information
- Journal Title
- Nuclear Instruments and Methods in Physics Research. Section A, Accelerators, Spectrometers, Detectors and Associated Equipment
- Journal Volume
- 591
- Journal Issue
- 1
- Journal Page Range
- p. 92-97
- ISSN
- 0168-9002
- CODEN
- NIMAER
Conference
- Title
- 9. international workshop on radiation imaging detectors
- Dates
- 22-26 Jul 2007
- Place
- Erlangen (Germany)
INIS
- Country of Publication
- Netherlands
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 40033949
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Resource subtype / Literary indicator
- Conference
- Descriptors DEI
- ENERGY RESOLUTION; ION SPECTROSCOPY; JET TOKAMAK; LIGHT IONS; NEUTRAL PARTICLE ANALYZERS; NEUTRAL PARTICLES; NEUTRONS; PHOSPHORS; PHOTOMULTIPLIERS; PHOTONS; PLASMA; SCINTILLATION COUNTERS; SENSITIVITY; SI SEMICONDUCTOR DETECTORS; SIGNALS
- Descriptors DEC
- BARYONS; BOSONS; CHARGED PARTICLES; CLOSED PLASMA DEVICES; ELEMENTARY PARTICLES; FERMIONS; HADRONS; IONS; MASSLESS PARTICLES; MEASURING INSTRUMENTS; NUCLEONS; PHOTOTUBES; RADIATION DETECTORS; RESOLUTION; SEMICONDUCTOR DETECTORS; SPECTROMETERS; SPECTROSCOPY; THERMONUCLEAR DEVICES; TOKAMAK DEVICES
Optional Information
- Copyright
- Copyright (c) 2008 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.