Description

Arc-modulated radiation therapy (AMRT) is a novel rotational intensity-modulated radiation therapy (IMRT) technique developed for a clinical linear accelerator that aims to deliver highly conformal radiation treatment using just one arc of gantry rotation. Compared to fixed-gantry IMRT and the multiple-arc intensity-modulated arc therapy (IMAT) techniques, AMRT promises the same treatment quality with a single-arc delivery. In this paper, we present a treatment planning scheme for AMRT, which addresses the challenges in inverse planning, leaf sequencing and dose calculation. The feasibility and performance of this AMRT treatment planning scheme have been verified with multiple clinical cases of various sites on Varian linear accelerators.

Availability note (English)

Available from http://dx.doi.org/10.1088/0031-9155/53/22/002

Additional details

Identifiers

DOI
10.1088/0031-9155/53/22/002;
PII
S0031-9155(08)82278-6;

Publishing Information

Journal Title
Physics in Medicine and Biology
Journal Volume
53
Journal Issue
22
Journal Page Range
p. 6291-6303
ISSN
0031-9155
CODEN
PHMBA7

INIS

Country of Publication
United Kingdom
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
41004152
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
LINEAR ACCELERATORS; PERFORMANCE; PLANNING; RADIOTHERAPY
Descriptors DEC
ACCELERATORS; MEDICINE; NUCLEAR MEDICINE; RADIOLOGY; THERAPY