Published 1992
| Version v1
Report
Open
TL-based dosemeter for measurement of low level photon radiation
Description
Results from the calibration of a TL-based personal dosemeter for measurements of low levels of external photon radiation are presented. The lower limit of detection is 2 to 3 μGy, depending on the photon energy. In the energy range 40 to 180 keV the dosemeter will also provide information concerning the effective energy of the photon field. The uncertainty in the dose measurements is estimated to 10% for the energy range 40 to 180 keV, and the same for natural background radiation/Cs-fallout. For photon radiation of unknown energy in the range 180 keV to 1.24 MeV, the uncertainty due to the energy response is approx. 30%. 5 refs., 5 figs., 3 tabs
Availability note (English)
MF available from INIS under the Report Number.Files
24007568.pdf
Files
(180.2 kB)
| Name | Size | Download all |
|---|---|---|
|
md5:dce032e186e9a3f50585f7fe1713309c
|
180.2 kB | Preview Download |
Additional details
Additional titles
- Subtitle (English)
- Description and calibration
- Original title (Norwegian)
- TL-basert dosimeter for maaling av lave eksterndosenivaaer
- Original subtitle (Norwegian)
- Beskrivelse og kalibrering.
Publishing Information
- Imprint Pagination
- 14 p.
- ISSN
- 0803-852X
- Report number
- SIS--1992:7
INIS
- Country of Publication
- Norway
- Country of Input or Organization
- Norway
- INIS RN
- 24007568
- Subject category
- S46: INSTRUMENTATION RELATED TO NUCLEAR SCIENCE AND TECHNOLOGY;
- Descriptors DEI
- CALCIUM FLUORIDES; CALIBRATION; LOW DOSE IRRADIATION; PHOTONS; THERMOLUMINESCENT DOSEMETERS; THERMOLUMINESCENT DOSIMETRY
- Descriptors DEC
- ALKALINE EARTH METAL COMPOUNDS; BOSONS; CALCIUM COMPOUNDS; CALCIUM HALIDES; DOSEMETERS; DOSIMETRY; ELEMENTARY PARTICLES; FLUORIDES; FLUORINE COMPOUNDS; HALIDES; HALOGEN COMPOUNDS; IRRADIATION; LUMINESCENT DOSEMETERS; MASSLESS PARTICLES; MEASURING INSTRUMENTS