Published 1992 | Version v1
Report Open

TL-based dosemeter for measurement of low level photon radiation

Description

Results from the calibration of a TL-based personal dosemeter for measurements of low levels of external photon radiation are presented. The lower limit of detection is 2 to 3 μGy, depending on the photon energy. In the energy range 40 to 180 keV the dosemeter will also provide information concerning the effective energy of the photon field. The uncertainty in the dose measurements is estimated to 10% for the energy range 40 to 180 keV, and the same for natural background radiation/Cs-fallout. For photon radiation of unknown energy in the range 180 keV to 1.24 MeV, the uncertainty due to the energy response is approx. 30%. 5 refs., 5 figs., 3 tabs

Availability note (English)

MF available from INIS under the Report Number.

Files

24007568.pdf

Files (180.2 kB)

Name Size Download all
md5:dce032e186e9a3f50585f7fe1713309c
180.2 kB Preview Download

Additional details

Additional titles

Subtitle (English)
Description and calibration
Original title (Norwegian)
TL-basert dosimeter for maaling av lave eksterndosenivaaer
Original subtitle (Norwegian)
Beskrivelse og kalibrering.

Publishing Information

Imprint Pagination
14 p.
ISSN
0803-852X
Report number
SIS--1992:7