Published February 2004 | Version v1
Journal article

Evaluation of a 2D diode array for IMRT quality assurance

Description

Background and purpose: The QA of intensity modulated radiotherapy (IMRT) dosimetry is a laborious task. The goal of this work is to evaluate the dosimetric characteristics of a new 2D diode array (MapCheck from Sun Nuclear Corporation, Melbourne, Florida) and assess the role it can play in routine IMRT QA. Material and methods: Fundamental properties of the MapCheck such as reproducibility, linearity and temperature dependence are studied for high-energy photon beams. The accuracy of the correction for difference of diode sensitivity is also assessed. The diode array is benchmarked against film and ion chambers for conventional and IMRT treatments. The MapCheck sensitivity to multileaf collimator position errors is determined. Results: The diode array response is linear with dose up to 295 cGy. All diodes are calibrated to within ±1% of each other, and mostly within ±0.5%. The MapCheck readings are reproducible to within a maximum SD of ±0.15%. A temperature dependence of 0.57%/ degree sign C was noted and should be taken into account for absolute dosimetric measurement. Clinical performance of the MapCheck for relative and absolute dosimetry is demonstrated with seven beam (6 MV) head and neck IMRT plans, and compares well with film and ion chamber measurements. Comparison to calculated dose maps demonstrates that the planning system model underestimates the dose gradients in the penumbra region. Conclusions: The MapCheck offers the dosimetric characteristics required for performing both relative and absolute dose measurements. Its use in the clinic can simplify and reduce the IMRT QA workload

Additional details

Identifiers

DOI
10.1016/j.radonc.2003.10.014;
PII
S0167814003004171;

Publishing Information

Journal Title
Radiotherapy and Oncology
Journal Volume
70
Journal Issue
2
Journal Page Range
p. 199-206
ISSN
0167-8140
CODEN
RAONDT

INIS

Country of Publication
Ireland
Country of Input or Organization
International Atomic Energy Agency (IAEA)
INIS RN
35052549
Subject category
S62: RADIOLOGY AND NUCLEAR MEDICINE;
Descriptors DEI
ACCURACY; COLLIMATORS; DOSIMETRY; ERRORS; IONIZATION CHAMBERS; PHOTON BEAMS; PLANNING; QUALITY ASSURANCE; RADIATION DOSES; RADIOTHERAPY
Descriptors DEC
BEAMS; DOSES; MEASURING INSTRUMENTS; MEDICINE; NUCLEAR MEDICINE; RADIATION DETECTORS; RADIOLOGY; THERAPY

Optional Information

Copyright
Copyright (c) 2003 Elsevier Science B.V., Amsterdam, The Netherlands, All rights reserved.