Published 2011 | Version v1
Journal article

Different versions of the right answer: The importance of measurement uncertainty in radiation dosimetry

  • 1. Health Protection Agency, Chilton, Didcot, Oxon OX11 0RQ (United Kingdom)
  • 2. Babcock Nuclear Limited, Berkeley Approved Dosimetry Service, A11, Berkeley Centre, Berkeley, Glos GL13 9FB (United Kingdom)

Description

The performance of radiation dosemeters that are issued by approved individual monitoring services generally meet international standards, with typical results within a few tens of per cent of the reference value. Experienced dosimetry practitioners will understand the uncertainties and treat monitoring results with due caution. However, where different technologies (for example, where passive and electronic dosemeters) are used side by side, apparent disagreements can arise. These apparent disagreements between different systems can be significant, and dosimetrists must be prepared to help in addressing the issues that result. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncq298

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
144
Journal Issue
1-4
Journal Page Range
p. 62-66
ISSN
0144-8420

Conference

Title
European conference on individual monitoring of ionising radiation
Acronym
IM2010
Dates
8-12 Mar 2010
Place
Athens (Greece)

Optional Information

Notes
7 refs