Published 1986 | Version v1
Book

Design and dosimetry of an eye plaque containing I-125 seeds: An improved dose distribution

Description

To treat intraocular tumors, a temporarily implanted eye plaque, containing 24 I-125 seeds (3M model 6711), was fabricated from 0.6-mm-thick lead disk with a 1.5-cm diameter. The I-125 seeds were distributed in a particular geometric pattern to average the dose anisotropy of each individual seed. Water phantom measurements made with TLD chips (LiF) and film over the approximate depth range of 1-25 mm were compared with treatment planning computer calculations (Capintec RT-108). Data indicate that the specified geometry produces a dose distribution delivering a tumor dose of 10,000 rad to the tumor apex (7 mm) without exceeding a sclera dose (1 mm) of 40,000 rad. Information regarding fabrication, dosimetry, and radiation safety is presented

Additional details

Publishing Information

Publisher
Radiological Society of North America Inc.
Imprint Place
Oak Brook, IL (USA)
Imprint Title
The 72nd scientific assembly and annual meeting of the Radiological Society of North America (Abstracts)
Journal Page Range
p. 249.

Conference

Title
72. scientific assembly and annual meeting of RSNA.
Dates
30 Nov - 5 Dec 1986.
Place
Chicago, IL (USA).