Published 2022 | Version v1
Journal article

Surface finishing and shape effects on corrosion resistance of Ti-6Al-4V alloy in the simulated body fluid

  • 1. Universidade Federal de São João Del-Rei (UFSJ), MG (Brazil)
  • 2. Universidade do Estado do Rio de Janeiro (UERJ), Nova Friburgo, RJ (Brazil)

Description

Pure titanium and various alloys, such as Ti-6Al-4V, are widely used as biomaterials. In this application, surface finish conditions, topography, and surface reactivity determine excellent cellular adhesion and osseointegration characteristics. The study evaluated the influence of four surface finishing treatments (sanding # 120, 600, 1200, and polished with 3 μm alumina paste) and three shapes (concave, convex, and flat) on the corrosion resistance of Ti-6Al-4V at 37 °C in simulated body fluid (SBF) through potentiodynamic polarization and electrochemical impedance spectroscopy. The SBF's ionic composition allowed the formation of a stable passive layer with a low presence of pores on the surface. In addition, the combination of polished surface and convex shape showed the best electrochemical passive behavior. (author)

Additional details

Publishing Information

Journal Title
Materials Research (Sao Carlos, Online)
Journal Volume
25
Journal Page Range
11 p.
ISSN
1980-5373