A digital multi-channel spectrometer with temperature peak-shift correction function has been introduced in this paper. The digital signal processing circuit based on FPGA has been designed in our development. The energy spectrum shift of the LaBr3 detector at different ambient temperature has been studied and a temperature compensation algorithm has been designed. The test shows that the developed digital multi-channel spectrometer has good energy resolution and energy linearity, and the temperature peak-shift correction method is effective for LaBr3 detector. (authors)

Additional details

Publishing Information

Journal Title
Nuclear Electronics and Detection Technology
Journal Volume
37
Journal Issue
9
Journal Page Range
p. 946-951
ISSN
0258-0934

Optional Information

Notes
10 figs., 13 refs.