Published 2013 | Version v1
Journal article

Suspension criteria for dual energy X ray absorptiometry

Creators

  • 1. Medical Physics and Radiation Engineering, Canberra Hospital, Garran, ACT 2605 (Australia)

Description

The use of dual-energy X-ray absorptiometry (DXA) units primarily for the assessment of fracture risk and in the diagnosis of osteoporosis is ubiquitous in Europe and ever-expanding in its implementation worldwide. DXA is known for its reported low radiation dose and precision in the determination of bone mineral density. However, the use of simple suspension criteria, as proposed in the new EC report RP-162, will identify units that are unfit for useful and safe diagnosis. Such suspension levels, however, are not a substitute for regular maintenance, quality control testing and optimisation of clinical outcomes. (authors)

Availability note (English)

Available from doi: http://dx.doi.org/10.1093/rpd/ncs298

Additional details

Identifiers

Publishing Information

Journal Title
Radiation Protection Dosimetry
Journal Volume
153
Journal Issue
2
Journal Page Range
p. 227-229
ISSN
0144-8420

Optional Information

Notes
15 refs