Image reconstruction algorithm for optically stimulated luminescence 2D dosimetry using laser-scanned Al2O3:C and Al2O3:C,Mg films
Creators
- 1. Physics Department, Oklahoma State University, Stillwater, OK 74078 (United States)
- 2. Department of Radiation Oncology, The University of Oklahoma Health Sciences Center, Oklahoma City, OK 73104 (United States)
Description
The objective of this work was to develop an image reconstruction algorithm for 2D dosimetry using Al2O3:C and Al2O3:C,Mg optically stimulated luminescence (OSL) films imaged using a laser scanning system. The algorithm takes into account parameters associated with detector properties and the readout system. Pieces of Al2O3:C films (∼8 mm × 8 mm × 125 µ m) were irradiated and used to simulate dose distributions with extreme dose gradients (zero and non-zero dose regions). The OSLD film pieces were scanned using a custom-built laser-scanning OSL reader and the data obtained were used to develop and demonstrate a dose reconstruction algorithm. The algorithm includes corrections for: (a) galvo hysteresis, (b) photomultiplier tube (PMT) linearity, (c) phosphorescence, (d) 'pixel bleeding' caused by the 35 ms luminescence lifetime of F-centers in Al2O3, (e) geometrical distortion inherent to Galvo scanning system, and (f) position dependence of the light collection efficiency. The algorithm was also applied to 6.0 cm × 6.0 cm × 125 μm or 10.0 cm × 10.0 cm × 125 µ m Al2O3:C and Al2O3:C,Mg films exposed to megavoltage x-rays (6 MV) and 12C beams (430 MeV u−1). The results obtained using pieces of irradiated films show the ability of the image reconstruction algorithm to correct for pixel bleeding even in the presence of extremely sharp dose gradients. Corrections for geometric distortion and position dependence of light collection efficiency were shown to minimize characteristic limitations of this system design. We also exemplify the application of the algorithm to more clinically relevant 6 MV x-ray beam and a 12C pencil beam, demonstrating the potential for small field dosimetry. The image reconstruction algorithm described here provides the foundation for laser-scanned OSL applied to 2D dosimetry. (paper)
Availability note (English)
Available from http://dx.doi.org/10.1088/0031-9155/61/20/7484Additional details
Identifiers
Publishing Information
- Journal Title
- Physics in Medicine and Biology
- Journal Volume
- 61
- Journal Issue
- 20
- Journal Page Range
- p. 7484-7506
- ISSN
- 0031-9155
- CODEN
- PHMBA7
INIS
- Country of Publication
- United Kingdom
- Country of Input or Organization
- International Atomic Energy Agency (IAEA)
- INIS RN
- 49094708
- Subject category
- S62: RADIOLOGY AND NUCLEAR MEDICINE; S61: RADIATION PROTECTION AND DOSIMETRY;
- Descriptors DEI
- ALGORITHMS; ALUMINIUM OXIDES; BEAMS; CARBON 12; CORRECTIONS; DESIGN; DOSIMETRY; EFFICIENCY; HYSTERESIS; IMAGE PROCESSING; IMAGES; IRRADIATION; PHOSPHORESCENCE; PHOTOMULTIPLIERS; RADIATION DOSE DISTRIBUTIONS; RADIATION DOSES; READOUT SYSTEMS; SPACE DEPENDENCE
- Descriptors DEC
- ALUMINIUM COMPOUNDS; CARBON ISOTOPES; CHALCOGENIDES; DOSES; EMISSION; EVEN-EVEN NUCLEI; ISOTOPES; LIGHT NUCLEI; LUMINESCENCE; MATHEMATICAL LOGIC; NUCLEI; OXIDES; OXYGEN COMPOUNDS; PHOTON EMISSION; PHOTOTUBES; PROCESSING; STABLE ISOTOPES