Description

A variable-size 'box' method is described which optimizes the precision with which integrated intensities of neutron Bragg reflections are measured with an Anger camera. The consequent improvements in the least-square-fitted crystallographic structural parameters and R factors in test experiments on MnF2 and Cr3Si are substantial. (orig.)