Published August 1988 | Version v1
Journal article

A gas flow proportional detection system for the determination of boron in aerosol samples and other sample types

Creators

  • 1. Lowell Univ., MA (USA). Dept. of Physics

Description

The intent of this work was to develop a detector which would exhibit high detection efficiency for the 10B(n,α)7Li reaction products emitted when aerosol samples on high efficiency filters were irradiated in reasonably intense thermal neutron fields so that acceptable results could be obtained through the nondstructive technique in reasonable analysis times. Testing and evaluation demonstrated that the lucite-bodied internal gas proportional detector can be used effectively to analyze particulate aerosols for boron-10 content in short analysis times, given a reasonably high thermal neutron fluence rate. A maximum sensitivity of 1.73 x 10-6 cpm per (n cm-2 s-1) per nanogram of natural boron was determined for the detector system described. (author) 14 refs.; 8 figs.; 4 tabs

Additional details

Publishing Information

Journal Title
Journal of Radioanalytical and Nuclear Chemistry
Journal Volume
123
Journal Issue
2
Series
J. Radioanal. Nucl. Chem. ;Articles.
Journal Page Range
491-515
ISSN
0236-5731
CODEN
JRNCD

Conference

Title
International topical conference on methods and applications of radioanalytical chemistry.
Dates
5-10 Apr 1987.
Place
Kona, HI (USA).